Medical Oxygen Purity: What the Numbers Mean
Oxygen purity is quoted as a number — 99.5 percent, 99.99 percent, 93 percent — and for a buyer those figures are close to interchangeable and almost entirely uninformative. The number matters far less than four things around it: what it is measured against, where it is measured, what else is in the gas, and who verified it.
What purity means, and what it does not
Medical oxygen is not a compound with a single fixed composition. It is a gas mixture held to a specification: a minimum oxygen concentration, and maximum limits on everything else in it. A sheet saying 99.5 percent states the oxygen floor and, unless the other limits are also given, tells you nothing at all about the remaining half of the gas.
- Oxygen concentration — the headline figure, and the least interesting one on the sheet.
- Carbon monoxide — a contaminant with no useful threshold; the limit is a hard ceiling and the measurement is critical.
- Carbon dioxide — limited for both respiratory and equipment reasons, and relevant wherever gas passes through a regulator or an absorber.
- Nitrogen — the balance gas in a cylinder supply, and the major component in an ambient-air system.
- Total hydrocarbons, including methane and heavier fractions — relevant to contamination carried in from compressed air or from a compressor without adequate separation.
- Water content, expressed as dew point — relevant to downstream equipment, and a common cause of behaviour that gets attributed to something else.
- Particulate and specific contaminants such as chlorine or ammonia — discussed less often and occasionally more important than the headline figure.
That list is the reason a bare percentage is not enough. A gas at 99.9 percent oxygen with carbon monoxide sitting at the limit is a different proposition from one at 99.5 percent with carbon monoxide undetectable, and a sheet quoting only the first number hides the difference completely.
Where the numbers come from
Different supply routes have entirely different profiles, and the figures a buyer encounters depend mostly on which route is being quoted.
- Cylinder supply, medical grade. Typically 99.5 percent or higher, produced by compressing and purifying air, with a certificate of analysis per cylinder or per batch. The weakness is that quality depends on the fill and on handling.
- Bulk liquid oxygen from a road tanker. Produced at large-scale plants and delivered at high purity. The risk here is not the purity but the interface — how the tanker coupling and the site storage are managed.
- On-site PSA generation. Output is typically 90 to 95 percent oxygen with the balance nitrogen, because the process separates nitrogen out of air and the remainder leaves the system. This is a design characteristic rather than a quality level, and it is predictable and repeatable.
- Membrane and electrochemical generation. Similar principle, different trade-offs around recovery and purity at low flow.
- Ambient air. Around 21 percent oxygen, used where pressure rather than concentration is the point of the exercise.
Why 93 percent appears so often
One figure confuses buyers more than any other, and it appears on a great deal of equipment. It is usually explained as a measurement artefact: the reading is taken at the outlet, where a small amount of room air is entrained into the flow, so the figure at the source is higher. That explanation is generally correct for a well-designed system, and it is also why the measurement point has to be specified precisely.
The practical implication is that two devices can both be correctly specified and still show different numbers at the same test point. Compare at the same point, with the same method, before drawing any conclusion.
The measurement question
- Where is the sample taken — the generator outlet, the pipeline, the terminal unit, or the point of use?
- What method is used: a paramagnetic or electrochemical analyser for concentration, a specific cell for carbon monoxide, and a dew point measurement for water?
- How often is the instrument calibrated, against what reference, and by whom?
- Is the measurement continuous with an alarm at a defined threshold, or a periodic spot check?
- Who records the results, and for how long are they retained?
A purity specification nobody measures is a wish rather than a specification. Continuous monitoring with an alarm is the difference between knowing and assuming, and it is a modest addition to a system that people generally consider over-specified until something drifts.
Matching the number to the application
The practical use of a purity figure is to match a gas to an application and to set a sensible alarm threshold. The matching is straightforward once the requirement is written down.
- Where a defined pharmacopoeia grade is required for a medical gas system, the specification is the standard itself and the supplier must demonstrate compliance against it, not against a number printed on a brochure.
- Where the requirement is industrial, food-grade or a laboratory application, the thresholds and the applicable standards differ — check which applies rather than assuming the medical figures are the stricter ones.
- Where the requirement is a chamber that enriches its own interior, the concentration that matters is the one inside the chamber during a session, and the supplier should state it as such.
- Where the requirement is a concentrator used at the bedside, what matters is the concentration at the patient end of the cannula, not at the machine, and the difference is larger than most buyers expect.
Questions to ask a supplier
- What is the guaranteed oxygen concentration, and at what point in the system is it guaranteed?
- What is the full specification — carbon monoxide, carbon dioxide, hydrocarbons, water content — rather than only the headline figure?
- Which standard or pharmacopoeia specification does the product conform to, and for which markets?
- How is it measured, how often is the instrument calibrated, and is there a continuous alarm?
- Is the certificate of analysis provided per cylinder, per batch, or per installation?
- Does the concentration hold at the minimum flow setting as well as at the maximum?
So to answer the question in the title: the purity percentage is the least informative part of the specification. What you actually need is where it is measured, the full impurity specification, the standard it is written against, and evidence that somebody checks it. A supplier who can answer those four has told you something useful; one who quotes a percentage on its own has not.
If you are specifying a gas supply, send us the standard that applies in your market and the flow, and we will work out what the measurement and monitoring should cover.